4 Preoperative Checkout
M1132382
4-3
Before every patient
Note
This check does not need to be done before the first case of the day if
the
“Every day before your first patient”
check was done.
Check that necessary emergency equipment is available and in good
condition.
Check the vaporizer installation:
•
The top of each vaporizer is horizontal (not on crooked).
•
Each vaporizer is locked and cannot be removed.
•
The alarms and indicators operate correctly (Tec 6 series
vaporizers).
•
More than one vaporizer cannot be turned on at the same time.
•
All vaporizers are full.
Do the Breathing system tests.
Set the appropriate controls and alarm limits for the case.