2 Theory of Operation
1006-0836-000 02/03
2-19
2.5 Mechanical Subsystems
Refer to:
•
Figure 9-3,
“Aestiva 7100 anesthesia machine pneumatic diagram,”
•
Figure 9-4,
“Aespire 7100 anesthesia machine pneumatic diagram,”
in section 9 for the complete pneumatic/mechanical subsystem.
The mechanical subsystem includes:
Pneumatic Engine
•
Drive gas inlet filter
•
Supply gas pressure regulator
•
Inspiratory flow control valve
•
PEEP valve; PEEP safety valve; Pressure sense switch
•
Mechanical Overpressure Valve (MOPV)
•
Bleed resistor
•
Free breathing valve
Exhalation valve
Breathing circuit flow sensors
Bellows assembly
2.5.1 Supply Gas
Supply gas (can be selected from O
2
or Air) is supplied from the anesthesia machine
at a pressure of 241 to 690 kPa (35 to 100 psi). This supply gas is filtered through the
2-micron filter to remove any minute particles of contaminate. The filter does not
significantly lower the output pressure on the downstream side of the filter.
2-MICRON
INLINE INLET
FILTER
SUPPLY GAS
35-100 psi
Coarse side up
Smooth side up (
under plate
)
Aestiva Engine
Aespire Engine