5 Troubleshooting
1006-0836-000 02/03
5-17
5.4.3 Inaccurate volume
ventilation troubleshooting
* Water in harness is indicative of a leak in the tubing, connectors, or MIA/VMB pressure transducer.
Be sure to leak check if it is determined that water is present within the harness.
AB.43.147
Correct kinks or
leaks, replace
harness if
necessary.
*
Start
Zero Flow Sensor
Transducers and
Replace both
flow sensors
(whole cartridge)
Perform a
Pressure
Sensitivity
Calibration
Do Bit Counts on either
channel vary by more
than 10 from the initial
zero count?
Inspect for leaks
or water in
pneumatic lines
between flow sensor
and MIA/VMB.
Problem
Continues?
Yes
No
Sensor Problem:
Examine for defect or water
plug in lines. Ensure
customer is periodically
emptying integrated circle
module water trap
Water, leaks
or kinks?
Yes
Switch insp. and exp.
pneumatic connections
and repeat Pressure
Sensitivity Cal
Yes
No
Problem
moves to
other sensor
channel?
Replace
Harness
Yes
No
Inspect Breathing
Circuit check
valves: replace
seat and disk as
necessary
No
Problem
Continues?
No
Pressure Sensitivity
Cal Error:
Yes
Problem
Continues?
Yes
Check Valve Problem:
MIA/VMB Board Evaluation
No
Problem with
MIA/VMB indicated?
Yes
Replace
MIA/VMB
Replace
Harness
Perform
Calibrations and
Re-evaluate
Perform
Calibrations and
Re-evaluate
No
Refer to Section 5.4.4.A, "MIA board evaluation (Aestiva machine)."
or
Refer to Section 5.4.5.B, "VMB board evaluation (Aespire machine)."