7100 Anesthesia Ventilator
7-22
02/03 1006-0836-000
To replace
the power cord/harness
1. Remove the top shroud (
A
) (refer to Aestiva Machine Service Manual).
2. Remove the harness connector (
B
) from mounting bracket.
3. Disconnect the ground wires (
C
).
4. Disconnect the wires from the Tec 6 outlet (
D
).
5. Disconnect the power cord from the control module (
E
).
6.
For machines with Display Arm
, disassemble arm to gain access to
routing for power cord
•
Remove cover at end of arm.
•
Remove top cover of arm.
•
Remove front and back plates from shoulder.
•
Loosen shoulder mounting screws enough to form gap at top large
enough to pass the power plug connector.
For machines with Folding Mount
, remove the side panel.
7. Pull the power cord into the electrical enclosure.
8. Reassemble in reverse order.
9. Apply
RTV
where the cord enters the machine area from the top.
B
D
B
A
D
C
E
RTV
Folding
Mount
A R M