4 Tests and Calibration
1006-0836-000 02/03
4-3
4.2 Service Mode
The
Service Mode
is used to test, calibrate, or troubleshoot ventilator related
components in the anesthesia machine.
There are two ways to enter the service mode:
•
If the machine is turned off, push and hold in the adjustment knob
while setting the system switch to On. Hold the adjustment knob
pushed in until the “Service Confirmation” menu appears. Use the
adjustment knob to highlight “Service Mode”, then push the
adjustment knob to confirm the selection.
•
If the machine is already in normal operation, set the Bag/Vent switch
to Bag. Then, press the V
T /
Pinsp, the PEEP, and the menu switches at
the same time to reset the software (powerup). Push and hold the
adjustment knob until the “Service Confirmation” menu appears.
Note: Volume alarms must be in standby.
From this menu you can:
•
Go to the Service Modes
•
Return to Normal Operations
Service Modes Menu
The
Service Modes
main menu displays the service tests you can select.
The selectable service tests are displayed in categorical order. But you can
select the service test from this menu in any order. The following sections in this
manual are sequenced in the order that they appear on the screen.