4 Tests and Calibration
1006-0836-000 02/03
4-5
4.4 Alarm Log
The
Alarm Log
displays up to 20 of the most current alarm messages that have
been logged. Each log entry shows:
•
Bootup Count
number (the bootup count is incremented each time
the machine is turned on).
•
Time (ms)
— the time in milliseconds since bootup when the latest
alarm condition occurred.
•
Alarm
message associated with the particular alarm condition.
•
# Times
— the number of times that the specific alarm condition has
occurred during the noted bootup count.
The bottom-left corner of the screen displays additional information:
•
Bootup Count Last Cleared
•
Current Bootup Count
Scroll Data
To view all the log entries, select “Scroll Data.”
Then, use the control knob to scroll through the entries.
To exit Scroll Data, press the control knob.
Clear Alarm Log
To clear the alarm log, select “Clear Alarm Log.” The system asks you to confirm
that you want to clear the alarm log.
Remarks
After you clear the alarm log:
•
the “Bootup Count Last Cleared” number will be reset to the “Current
Bootup Count” number.
•
the menu will show the message “No entries in log!”.