7 Repair Procedures
1006-0836-000 02/03
7-17
7.3.7 Manifold and
plate assembly
Manifold plate gasket
(A)
1. Ensure that gasket is properly positioned.
2. Carefully install plate onto manifold making sure not to disturb the gasket.
3. Start all screws first.
4. Then, torque to 1.7 N-m (15 lb-in) using sequence shown.
b
a
d
c
e
A
Aestiva machine
b
a
d
c
e
f
g
A
Aespire machine