4 Tests and Calibration
1006-0836-000 02/03
4-21
7. To start calibration select “Start Pressure Sensitivity Cal”.
•
Adjust the APL and fresh gas flow until the real-time pressure reads a
stable 10 cm H
2
O
— after 5 seconds, select “Save Value”.
•
Adjust the APL and fresh gas flow until the real-time pressure reads a
stable 20 cm H
2
O
— after 5 seconds, select “Save Value”.
•
Adjust the APL and fresh gas flow until the real-time pressure reads a
stable 40 cm H
2
O
— after 5 seconds, select “Save Value”.
•
Adjust the APL and fresh gas flow until the real-time pressure reads a
stable 60 cm H
2
O
— after 5 seconds, select “Save Value”.