7 Repair Procedures
1006-0836-000 02/03
7-5
7.2.1 Inside the control
module
The control module consists of two enclosures: the rear enclosure (
A
) and the
front enclosure (
B
). To access components within each enclosure, separate
the two halves.
Opening the control module
1. Place the control module face down on a protected surface.
2. Loosen the screws (
C
) that hold the two enclosures together (one in each
corner).
3. Slightly separate the rear enclosure (
A
) from the front enclosure (
B
).
4. Swing the rear enclosure into an upright position. Lift up the edge that
houses the external connectors. Be careful not to stress the harnesses in
the back.
5. Disconnect the harnesses from the back of control board (
CB
) to separate
the two enclosures.
•
Battery (
D
)
•
Ground (
E
)
•
Power supply (
F
)
•
Fan (
G
)
w
w
w
w
Caution
When disassembling, remove the battery connector first. When
reassembling, attach the battery connector last. Power is supplied to the
control board as long as the battery is plugged in.
D
E
F
G
(CB)
B
B
A
C
Summary of Contents for 7100
Page 1: ...7100 Anesthesia Ventilator Technical Reference Manual ...
Page 102: ...Notes 5 22 02 03 1006 0836 000 ...
Page 144: ...Notes 8 16 02 03 1006 0836 000 ...
Page 156: ...9 12 02 03 1006 0836 000 Notes ...
Page 157: ......