2 Theory of Operation
1006-0836-000 02/03
2-23
2.5.7 Mechanical
Overpressure
Valve
The Mechanical Overpressure Valve (MOPV) is a mechanical valve that operates
regardless of electrical power. It functions as a third level of redundancy to the
ventilator's pressure limit control functions, supplying pressure relief at approximately
110 cm H
2
O.
2.5.8 Free
Breathing Valve
The ventilator is programmed to supply a specified number of breaths per minute to
the patient. If, in between one of these programmed cycles, the patient needs a breath
(spontaneous), the free breathing valve permits the patient to inhale. The free breathing
valve is closed on mechanical inspiration.
R
INSPIRATORY
CONTROL
VALVE
TEST
POINT
MECHANICAL
OVERPRESSURE
RELIEF
110 cm H2O
ATMOSPHERE
FREE BREATHING
CHECK VALVE
R
INSPIRATORY
CONTROL
VALVE
TEST
POINT
MECHANICAL
OVERPRESSURE
RELIEF
110 cm H2O
ATMOSPHERE
FREE BREATHING
CHECK VALVE