7100 Anesthesia Ventilator
4-28
02/03 1006-0836-000
4.10.4
Test Panel Switches
In the
Test Panel Switches
menu the software is set up to receive keyboard
button presses and rotary encoder turns.
Press each button on the panel and the control knob.
•
When a button is pressed, the icon on the screen next to the button
should be highlighted and filled with a checkmark.
•
When the button is released, the button icon should be reverse-
highlighted.
After testing all the buttons and the control knob, select “Test Encoder Knob
Turn” to test the encoder.
As you turn the encoder knob, verify that:
•
each click of the encoder in the clockwise direction increments the
clockwise total.
•
each click of the encoder in the counterclockwise direction
increments the counterclockwise total.
Remarks
If any of the select buttons test fails, replace the front panel keyboard
assembly.
If the encoder knob test fails, replace the rotary encoder assembly.
✓