7100 Anesthesia Ventilator
4-32
02/03 1006-0836-000
4.10.8
Test Serial Port
The
Test Serial Port
menu includes directions for two tests.
•
External Serial Port Testing
: The external test requires that you
jumper pins 6 and 13 of the serial connector.
•
Internal Serial Port Testing
: The internal test does not require any
setup; it only tests serial port related components on the control
board.
Each test routine sets up the serial port circuits so transmit data is echoed
directly back to the receive circuits. The test fails if the data sent out is not
equal to the data received.
Remarks
If only the internal test fails, replace the control board.
If both tests fail:
•
Check the harness connections between the control board and the
Serial board (SAB/SICB).
•
Check the ribbon cable between the SAB and the external connector.
•
Replace the SAB/SCIB.