DT9000
42
20171205-V2
7.7. The physical value assigned to 4 mA
(05. menu item)
Function
Defining the physical value that should be assigned to the
loop current value 4 mA, this can be any numerical value
within the range that can be displayed (-20000 - 20000).
[Default factory setting: 4.000]
For the purpose of scaling not only the physical value that
belongs to 4 mA, but also the physical value that belongs
to 20 mA must be set (the following menu item).
Sequence of operations
1. Enter the menu by typing the supervisor code.
The way the code should be type in is described in Chapter
7.1. Typing the code in.
You see on the display:
.
2. Keep stepping by pressing the
▲
button as long as you
see this menu item:
.
3. Enter the menu item by pressing the
ENTER
button.
4. You see
on the display, the arrow (
←
) is
blinking in the upper left corner. The arrow indicates that
the uppermost digit (the „half” digit) is being selected from
the 4 and half digits, and now it can be modified by
pressing the
▲
button.
5. Pressing the
▲
button you can switch the „half” digit on
or off, or you can type in the value 20.000.
•
= 14.000,
•
= 20.000
•
= 4.000
6. Pressing the
►
button you can select the next numerical
digit.
7. Pressing the
▲
button you can increase the value of the
blinking digit:
,
,
,
,
,
,
,
,
,
, etc.
8. When the last digit (right-side) is being selected, and you
press the
►
button, a blinking colon appears on the left
side:
. The colon indicates that the plus/minus
sign is being selected, and it can be modified by pressing
the
▲
button.
9. Pressing the
▲
button you can switch the minus sign on
or off.
•
= 4.000
•
= -4.000
Summary of Contents for DT9000
Page 1: ...DT9000 Intrinsically Safe Process Indicator Operating Instructions...
Page 65: ...DT9000 20171205 V2 65 10 2 Application example...
Page 85: ...DT9000 20171205 V2 85 10 9 ATEX Certification...
Page 86: ...DT9000 86 20171205 V2...
Page 87: ...DT9000 20171205 V2 87...
Page 88: ...DT9000 88 20171205 V2...
Page 89: ...DT9000 20171205 V2 89...
Page 90: ...DT9000 90 20171205 V2...
Page 91: ...DT9000 20171205 V2 91...
Page 92: ...DT9000 92 20171205 V2...
Page 93: ...DT9000 20171205 V2 93...
Page 94: ...DT9000 94 20171205 V2...
Page 95: ...DT9000 20171205 V2 95...
Page 96: ......