7.4. Setting up the limit outputs
(02. and 03. menu items)
Function
The instrument has two limit outputs. They are identical
with each other, and work independently.
Operation: the instrument keeps comparing the displayed
physical value with the (adjustable) limit value. It switches
the output ON depending whether the displayed value is
higher than the limit value. (There are other operating
modes too, in which it operates if the displayed value is
between or out of the two (adjustable) limit values.)
In Chapter
10.8. Limit outputs
in the Appendix you find a
detailed training material on the limit outputs.
Setting the operating
mode
Sequence of operations
1. Enter the menu with the supervisor code.
The way the code should be type in can be found in
Chapter
7.1. Typing the code in
. You see on the display:
.
2. Keep stepping by pressing the
▲
button as long as you
see this menu item:
.
3. Enter the menu item by pressing the
ENTER
button.
4. You see the mnemonic
on the display.
L1 (
L
imit 1) means: limit output No. 1.
SS (
S
etting
s
) means: settings. Here you can define the
operating mode for the limit output. Enter by pressing the
ENTER
button.
Summary of Contents for DT9000
Page 1: ...DT9000 Intrinsically Safe Process Indicator Operating Instructions...
Page 65: ...DT9000 20171205 V2 65 10 2 Application example...
Page 85: ...DT9000 20171205 V2 85 10 9 ATEX Certification...
Page 86: ...DT9000 86 20171205 V2...
Page 87: ...DT9000 20171205 V2 87...
Page 88: ...DT9000 88 20171205 V2...
Page 89: ...DT9000 20171205 V2 89...
Page 90: ...DT9000 90 20171205 V2...
Page 91: ...DT9000 20171205 V2 91...
Page 92: ...DT9000 92 20171205 V2...
Page 93: ...DT9000 20171205 V2 93...
Page 94: ...DT9000 94 20171205 V2...
Page 95: ...DT9000 20171205 V2 95...
Page 96: ......