10.7. Setting up the instrument (Example)
Exercise
There is a pressure transmitter with a measurement range
of 5-150 bar corresponds to 4-20 mA.
Let the settings of DT9000 be the following:
•
Number of decimals: 2.
•
The physical value assigned for the 4 mA current: 5.00
•
The physical value assigned for the 20 mA current:
150.00
•
The leader zeros are not displayed.
Enter the menu
Enter the menu by using a Supervisor Code.
The way the code should be typing in is described in
Chapter
7.1. Typing the code in.
You see on the display:
.
Setting the decimal point
1. Keep stepping by pressing the
▲
button as long as you
see this menu item:
.
2. Enter the menu item by pressing the
ENTER
button.
3. You see this
on the display. The mnemonic is
blinking.
4. Set the number of decimals to two by pressing the
▲
button:
.
5. Press the
MENU
button, to exit from the menu item, and
you see this:
.
Setting the physical
value assigned to 4 mA
1. Pressing the
▲
button keep stepping as long as you see
this menu item:
.
2. Enter the menu item by pressing the
ENTER
button.
3. Pressing the
▲
button you can increase the value of the
blinking number:
4. Pressing the
►
button you can select the next digit.
5. Set the value
.
6. Press the
MENU
button, to exit from the menu item, and
you see this:
.
Summary of Contents for DT9000
Page 1: ...DT9000 Intrinsically Safe Process Indicator Operating Instructions...
Page 65: ...DT9000 20171205 V2 65 10 2 Application example...
Page 85: ...DT9000 20171205 V2 85 10 9 ATEX Certification...
Page 86: ...DT9000 86 20171205 V2...
Page 87: ...DT9000 20171205 V2 87...
Page 88: ...DT9000 88 20171205 V2...
Page 89: ...DT9000 20171205 V2 89...
Page 90: ...DT9000 90 20171205 V2...
Page 91: ...DT9000 20171205 V2 91...
Page 92: ...DT9000 92 20171205 V2...
Page 93: ...DT9000 20171205 V2 93...
Page 94: ...DT9000 94 20171205 V2...
Page 95: ...DT9000 20171205 V2 95...
Page 96: ......