7.10. Display refresh time
(08. menu item)
Function
The instrument performs cca. 15 measurements in each
seconds. It’s too fast to see the changing of the
measurement value.
Here you can define the time periods by which the
instrument displays the new measurement values.
[Default factory setting: 0.5 second]
Sequence of operations
1. Enter the menu by pressing a supervisor code.
The way the code should be type in is described in Chapter
7.1. Type the code in.
You see on the display:
.
2. Keep stepping by pressing the
▲
button as long as you
see this menu item:
.
3. Enter the menu item by pressing the
ENTER
button.
4. You see this
on the display. It’s meaning: 0.5
seconds) The 0.5 numerical value is blinking.
5. Pressing the
▲
button select the time periods by which
the instrument should display the new measurement
values.
•
= once in each 0.5 seconds [Default factory
setting].
•
= once in each 1 second.
•
= once in each 1.5 seconds.
•
= once in each 2 seconds.
•
= once in each 0.1 seconds.
•
= once in each 0.3 seconds.
Summary of Contents for DT9000
Page 1: ...DT9000 Intrinsically Safe Process Indicator Operating Instructions...
Page 65: ...DT9000 20171205 V2 65 10 2 Application example...
Page 85: ...DT9000 20171205 V2 85 10 9 ATEX Certification...
Page 86: ...DT9000 86 20171205 V2...
Page 87: ...DT9000 20171205 V2 87...
Page 88: ...DT9000 88 20171205 V2...
Page 89: ...DT9000 20171205 V2 89...
Page 90: ...DT9000 90 20171205 V2...
Page 91: ...DT9000 20171205 V2 91...
Page 92: ...DT9000 92 20171205 V2...
Page 93: ...DT9000 20171205 V2 93...
Page 94: ...DT9000 94 20171205 V2...
Page 95: ...DT9000 20171205 V2 95...
Page 96: ......