Access is possible
by supervisor code only
Physical value assigned to 4 mA
The physical value to be assigned to the 4 mA loop current.
Any values can be defined for it within the minimum-
maximum interval that can be displayed . [Default factory
setting: 4 000]
Access is possible
by supervisor code only
Physical value assigned to 20 mA
The physical value to be assigned to the 20 mA loop
current. Any values can be defined for it within the
minimum-maximum interval that can be displayed. [Default
factory setting: 20 000]
Access is possible
by supervisor code only
Averaging number
(
A
veraging
N
umber)
The displayed measurement result is generated as the
average of several measurements. The number of
measurements can be defined here.
By increasing this number the display stability increases
too, but the signal settling time becomes lower.
[Default factory setting: 8]
Access is possible
by supervisor code only
Display refresh time
(
D
isplay
R
efresh)
The instrument performs cca. 15-16 measurements in each
seconds. It’s too fast to see the changing of the
measurement value.
Here you can define the time periods by which the
instrument displays the new measurement values.
[Default factory setting: 0.5 sec]
Test programmes sub-menu
(T
es
t)
•
Display testing (
T
est:
LCD
)
Checking if the display works correctly.
•
Output testing (
T
est:
Out
put)
Displaying and changing the status of the limit outputs,
independently of their operating modes.
Summary of Contents for DT9000
Page 1: ...DT9000 Intrinsically Safe Process Indicator Operating Instructions...
Page 65: ...DT9000 20171205 V2 65 10 2 Application example...
Page 85: ...DT9000 20171205 V2 85 10 9 ATEX Certification...
Page 86: ...DT9000 86 20171205 V2...
Page 87: ...DT9000 20171205 V2 87...
Page 88: ...DT9000 88 20171205 V2...
Page 89: ...DT9000 20171205 V2 89...
Page 90: ...DT9000 90 20171205 V2...
Page 91: ...DT9000 20171205 V2 91...
Page 92: ...DT9000 92 20171205 V2...
Page 93: ...DT9000 20171205 V2 93...
Page 94: ...DT9000 94 20171205 V2...
Page 95: ...DT9000 20171205 V2 95...
Page 96: ......