2.5 Main menu status icons
At certain times (e.g. after bump tests) a status icon
À
is displayed on the Main menu as shown in
below.
Figure 9: Status Icon
These status icons have the following meanings:
•
Action successful
•
Action failed
2.6 Replacing an empty gas cylinder
If a cylinder connected to the inlet is completely exhausted I-Test will recognised this and the icon
will be displayed on the screen.
It will be necessary to physically replace the cylinder and enter the new gas concentrations, Cylinder
Lot Number, and Cylinder Expiry Date via either I-Test front panel (see
and
) or
via I-Test Manager (see
and
).
i
If the cylinder was exhausted during a calibration the monitor may indicate a service fault
which must be cleared. Remove the monitor from I-Test and cancel all fault indications. Once
the cylinder has been replaced the monitor can then be reinserted and test can be repeated.
i
If I-Test detects an exhausted cylinder and the cylinder is physically changed, a new
Cylinder Lot number and Cylinder Expiry Date must be entered. If this process is not
followed, I-Test will not allow gas testing to continue and will indicate this on the screen.
2.7 I-Test Safe Shutdown
If it is required to remove power from I-Test either by removing the power lead or switching off via the
on/off push button then the safe shutdown process must be followed.
2012 21 Aug 10:25:00
À
•
Gas monitor needs calibration
•
Gas monitor unidentified/Sensor unidentified