Alter the following fields as required:
Allow
configuration:
I-Test can be configured to upload a new configuration file to the gas monitor
when inserted into I-Test.
This allows settings such as sensor alarm levels to be altered.
Select the
Allow configuration
option to enable this functionality.
For details on the settings for the configuration file see
Field Configuration
on
.
Speedy Bump:
Select this option to undertake a speedy bump on all sensors when the gas
monitor is inserted into I-Test. A speedy bump will test the gas monitor sensors
to the first alarm level.
Smart Bump:
(Gas-Pro only)
Select this option to undertake smart bump on all sensors when the gas
monitor is inserted into I-Test.
A smart bump will test that the gas monitor responds correctly to a specified
level of test gas within a predefined time window. These parameters are set in
the Inlet Configuration settings, see
.
i
Only one type of bump test can be selected, if speedy bump test is selected then smart
bump will be disabled and vice versa.
I-Test will not check calibration before undertaking a bump test and will continue to bump
test a gas monitor if calibration is due on a specific sensor (unless the Calibrate if Expired
option is selected).
Calibration:
I-Test offers thee calibration options dependant upon user requirements,
Standalone Calibration, Calibration after Bump Test Fail and Calibrate if
Expired.