and follow the on-screen instructions and the following installation screen will be shown.
Figure 12: ATEX or UL Region screen
X
I-Test Manager needs to be configured for the type of I-Test purchased, either ATEX region or
UL region compatible. This is due to the calculation I-Test will undertake to convert a volume
of flammable gas (%VOL) to a lower explosive limit value LEL, each region has a different
conversion factor.
X
Select the I-Test Manager region selection for either
ATEX
or
UL
as appropriate then click
Next
and follow the on-screen instructions (including the validation of any operating system
notifications) until the software is fully installed
i
Once I-Test Manager is configured to either ATEX or UL region, a matching I-Test must
only be used to calibrate or bump test that specific type of gas monitor. For example if
I-Test and I-Test Manager are configured for ATEX then UL calibrated gas monitors must
not be tested as they will be incorrectly tested or calibrated.