4.4.3.2 Bump Test Fail Calibration
This option should be selected if there is a requirement to calibrate a sensor, should it fail a smart bump test.
It should be noted that if a monitor passed a smart bump test but calibration on a sensor has
expired, at the end of the smart bump test this sensor will be calibrated.
i
If Calibration and Smart Bump are selected together I-Test will not perform a standalone
calibration upon insertion of the gas monitor (as described above) but will perform a
bump test fail calibration on any sensor that fails a smart bump test.
Only sensors that fail smart bump will be calibrated, this will occur even if the sensor was not due a calibration.
Bump Test Fail Calibration can be selected with Smart Bump, Allow configuration and download gas
monitor log files. Bump Test Fail calibration can
not
be selected with speedy bump.
i
If a monitor passed a smart bump test but calibration on a sensor has expired at the end
of the smart bump test, this sensor will be calibrated.
X
To select a Bump Test Fail Calibration, from the Test Sequence screen use the arrow keys to
navigate to the Calibration icon
then press
OK
to highlight the Calibration icon
À
. Then
use the arrow keys to navigate to the Smart Bump icon
then press
OK
to highlight the
Smart Bump icon
Á
.
X
A box will now appear around the icons indicating that Bump Test Fail Calibration function has
been selected and this test will be performed if a gas monitor is inserted into the I-Test.
i
To ensure a Bump Test Fail calibration is undertaken smart bump and calibration must
both be selected.
X
Press
to return to Test Settings screen and then press
again to return to main menu screen.
Note: Smart bump functionality is only applicable to Gas-Pro, a T4 I-Test test screen will not display
the smart bump icon
.
Figure 84: Bump Test Fail Calibration screen
13 May 2013 11:30:45
I-Test
À
Á
Contents
Exit
Unpacking
Set-up
Operating I-Test
I-Test
Manager
Configuring
I-Test
Warranty
Overview
Contacts
Appendices
Troubleshooting
Specification