Inlet Configuration
Inlet configuration allows the predefined cylinder profile, configured via the Manage Cylinders
), to be assigned to the gas inlet.
The screen shown below will enable the user to fully define the physical cylinder to be connected to
I-Test including actual gas values, Lot number and expiry date.
Figure 24: Inlet screen
Any alterations will cause the relevant field background to turn yellow.
Choose Cylinder:
Select the required cylinder profile from the drop down list.
This list will show all cylinder profiles previously saved to the database
Gases:
The gas or gases for a multi-gas cylinder profile will be displayed in the
Gases
list along with the standard value(s).
Actual Gas Value:
The actual gas value is the concentration of the gas in the physical cylinder to
be connected to I-Test. This should be read from the label on the cylinder
Click on each gas in turn, enter the actual gas values in the
Actual Value
field
and click
Update Value
. The entered value will be displayed in the Gases list.
Lot #:
This is the Lot number of the physical cylinder to be connected to I-Test. This
should be read from the label on the cylinder.
Expiry Date:
This is the expiry date of the physical cylinder to be connected to I-Test. This
should be read from the label on the cylinder.
Enter the correct date in the by clicking the calendar button to the right of the field.