cGas Detector Analog Transmitter - Operation Manual
The Calibration Kit and gases are available from the CETCI factory. Many gases,
but not all are carried in inventory. Check with any CETCI authorized distributor
for availability of specific gas types.
NOTE: Gas cylinders cannot be shipped
from Canada to other countries, including the USA.
11 MAINTENANCE
The cGas Detector transmitter requires virtually no maintenance other than
regular calibration of the sensor(s). There are no replaceable components
except the sensors.
The transmitter should be monitored for possible damaging conditions:
• The sensor vents should be kept free of dirt or soot build up.
•
If in a damp location, source of water should be shielded from contacting
the top of the transmitter.
• If located in a working area, the front of the transmitter should be kept
clear.
•
If painting is to be done in the same area as the transmitter, the
transmitter needs to be protected from over spray and the sensor vent
should be covered so as to not receive paint fumes. Paint fumes may
damage and / or reduce the life of the sensor.