34
© 2020 All rights reserved. Data subject to change without notice.
cGas Detector Analog Transmitter - Operation Manual
Rev. B | 2020.07
to change the value. Press ENTER to confirm and continue.
Alarm Hysteresis
>010 PPM CO2
Confirm?
>10 PPM
N
>Y
7.8 Change Units (oC or oF) of Temperature Readings
NOTE:
This menu item only applies if the cGas Detector has the -RHT option
installed.
You can change the factory configured temperature unit type from Celsius to
Fahrenheit (or vice versa) very easily.
From the main display, press ENTER.
Enter passcode 3032 using the UP button and ENTER button.
Enter Password
3032
Press ENTER to access the Calibrate Menu.
Calibrate Menu
Press Enter
In the Selected Channel menu, if Temperature is not displayed, press ENTER
and use the UP button to scroll to find Temperature. Press ENTER.
Selected Channel
CO2
Selected Channel
>Temperature
Use the UP button to find Temperature Unit. If you want to change the unit type,
press ENTER.
Temperature Unit
Celsius
Press the UP button to choose Fahrenheit press ENTER and ENTER to exit.
Temperature Unit
>Fahrenheit