7.9 Temperature and/or Relative Humidity Offset
NOTE:
This menu item only applies if the cGas Detector has the -RHT option
installed.
NOTE:
Depending on the configuration, the device will show the temperature
in either Celsius or Fahrenheit. The units can be changed at any time, refer to
Section
7.8 Change Units (oC or oF) of Temperature Readings
.
There is no way to calibrate a temperature or relative humidity sensor. However,
if you find that the temperature or humidity reading on the cGas Detector is
higher or lower than another device measuring the ambient temperature or
relative humidity, you can adjust the reading by setting an offset value so the
reading is more accurate. The Temperature offset value is a number of degrees
in either direction of 0 and the Humidity offset value is a percentage between 0
or 100)
From the main display, press ENTER.
Enter passcode 3032 using the UP button and ENTER button.
Enter Password
3032
Press ENTER to access the Calibrate Menu.
Calibrate Menu
Press Enter
In the Selected Channel menu, if Temperature (or Humidity) is not displayed,
press ENTER and use the UP button to scroll to find the item you are looking for.
Press ENTER.
Selected Channel
CO2
Selected Channel
>Temperature
Selected Channel
>Humidity
Use the UP button to find Temperature Adj (or Humidity Adj). If you want to
change the offset value, press ENTER. Change the offset value using the UP or
DOWN and ENTER buttons.