9.1.5 Gas Testing Frequency (Bump Testing)
For the purpose of safety in OHS applications, sensors should be gas tested
(bump tested) once every month to confirm response and alarm activation.
9.2 Calibrating the Internal Sensor(s)
Calibration has three processes - Set Calibration Gas Value, Span and Zero. The
Span and Zero calibrations can be done in any order.
9.2.1 Set Calibration Gas Value
Check to make sure that the calibration gas value configured in the device
matches the gas concentration of the calibration gas cylinder you are using.
This is especially important if you are not using the same Cal gas concentration
that was used previously to calibrate the device. How much gas the sensor
detects is directly related to the Cal gas setting and the actual concentration of
calibration gas used during calibration.
From the main display, press ENTER.
Enter passcode 3032 using the UP button and ENTER button.
Enter Password
3032
Press ENTER to access the Calibrate Menu.
Calibrate Menu
Press Enter
Selected Channel
CO2
Press the UP button to accept the displayed calibration gas value if it is correct.
Or press ENTER and change the value using the UP or DOWN buttons.
Calibration Gas
500 PPM CO2
Calibration Gas
>1000 PPM CO2
When finished pressed ENTER and then ENTER again to confirm and ENTER to
exit.
Confirm?
>1000 PPM
N
>Y