Size and Weight
Operator's Manual
9-9
9.9
Size and Weight
9.10
Factory Default Parameters
Table 9-11.
Size and Weight
Height (includes handle)
26.04 cm (10.25 in.)
Width
24.77 cm (9.75 in.)
Depth
27.94 cm (11 in.)
Weight
6.9 kg (15.4 lbs)
Table 9-12.
Factory Default Parameters (Patient Settings)
Mode
A/CMV
VT (Volume Control)
500 mL
i-time
1.0 sec
RR
12 b/min
Ptrig
2 cmH
2
O
3
Paw alarm
5 cmH
2
O
1
Paw alarm
40 cmH
2
O
3
MVI alarm
3 L/min
1
MVI alarm
9 L/min
1
O
2
Off
3
O
2
Off
PEEP/CPAP
Off
PS
Off
Alarm loudness
Level 6
Table 9-13.
Default Presets (Control Settings)
Control setting
Adult
Pediatric
Infant
Mode
VC A/CMV
PC A/CMV
PC A/CMV
VT
500 mL
N/A
N/A
PC
N/A
18 cmH
2
O
18 cmH
2
O
RR
12 b/min
20 b/min
20 b/min
Summary of Contents for Newport HT70
Page 1: ...Operator s Manual Newport HT70 Ventilator with Accessories Plus Model TM...
Page 8: ...Page Left Intentionally Blank viii Operator s Manual...
Page 12: ...Page Left Intentionally Blank xii Operator s Manual...
Page 24: ...Introduction Page Left Intentionally Blank 1 12 Operator s Manual...
Page 72: ...Navigating the Screens Page Left Intentionally Blank 4 24 Operator s Manual...
Page 88: ...Operating the Ventilator Page Left Intentionally Blank 5 16 Operator s Manual...
Page 100: ...Ventilator Alarms Page Left Intentionally Blank 6 12 Operator s Manual...
Page 108: ...Battery Operation Page Left Intentionally Blank 7 8 Operator s Manual...
Page 138: ...Explanation of Modes and Controls Page Left Intentionally Blank 10 6 Operator s Manual...
Page 140: ...Quick Check Procedure Check Off Page Left Intentionally Blank A 2 Operator s Manual...
Page 142: ...Foldout Drawings B 2 Page left intentionally blank...
Page 144: ...Foldout Drawings B 4 Page left intentionally blank...
Page 146: ...Foldout Drawings B 6 Page left intentionally blank...
Page 147: ......