Main Screen Navigation
Operator's Manual
4-9
4.5.3
Alarm Quickset
During ventilating (not standby) condition, when there are no active alarms violations,
Alarm
Quickset
will automatically set the alarm limits. Touch the
Alarm Quickset
button to enter the
Alarm Quickset screen, and then press
Accept
to activate or
Cancel
to return to the Alarms screen.
When activated, Alarm Quickset monitors settings for 30 seconds and then sets the alarms. If an
alarm occurs during the monitoring period, Quickset is canceled. During the 30-second period,
the touch screen will not respond unless an alarm occurs or the
Cancel
button is pressed. Alarm
Quickset will only activate when in ventilating condition.
4.6
Main Screen Navigation
The Main screen can be accessed in all domains and in standby or ventilating condition.
and
for ranges
and details for all parameters.
The ventilation parameters that are displayed on the Main screen are determined by the mode
and breath type that are selected with the
Mode/breath type
buttons along the left margin of the
touch screen. Select the mode and breath type first to see the Main screen parameters that need
adjustment.
Figure 4-6.
Main Screen
To set ventilation parameters (not available in basic domain):
1.
Touch a ventilation parameter button to activate (highlight) it.
2.
Use the up and down arrow panel buttons to adjust the setting.
3.
Press
Accept
to confirm changes or press
Cancel
to return to original settings.
Multiple parameters can be set before pressing
Accept
.
Summary of Contents for Newport HT70
Page 1: ...Operator s Manual Newport HT70 Ventilator with Accessories Plus Model TM...
Page 8: ...Page Left Intentionally Blank viii Operator s Manual...
Page 12: ...Page Left Intentionally Blank xii Operator s Manual...
Page 24: ...Introduction Page Left Intentionally Blank 1 12 Operator s Manual...
Page 72: ...Navigating the Screens Page Left Intentionally Blank 4 24 Operator s Manual...
Page 88: ...Operating the Ventilator Page Left Intentionally Blank 5 16 Operator s Manual...
Page 100: ...Ventilator Alarms Page Left Intentionally Blank 6 12 Operator s Manual...
Page 108: ...Battery Operation Page Left Intentionally Blank 7 8 Operator s Manual...
Page 138: ...Explanation of Modes and Controls Page Left Intentionally Blank 10 6 Operator s Manual...
Page 140: ...Quick Check Procedure Check Off Page Left Intentionally Blank A 2 Operator s Manual...
Page 142: ...Foldout Drawings B 2 Page left intentionally blank...
Page 144: ...Foldout Drawings B 4 Page left intentionally blank...
Page 146: ...Foldout Drawings B 6 Page left intentionally blank...
Page 147: ......