Automatic Alarms
Operator's Manual
6-7
6.5
Automatic Alarms
The following alarms are automatically set by the ventilator based on patient settings or device
condition. Violated alarms are indicated by an audible alarm, an alarm message displayed on the
touch screen, and the LEDs in the handle flashing.
+ only available on the HT70PM when the on-airway flow sensor is in use
6.5.1
High Baseline Pressure Alarm
The High Baseline Pressure alarm is activated by obstruction or high resistance to exhalation.
Check for obstructions to patient exhalation or for improper exhalation valve function. This could
be caused by aerosol medication buildup on the exhalation valve, an occluded filter, or incom-
plete exhalation due to auto-triggering.
6.5.2
Low Baseline Pressure Alarm
The Low Baseline Pressure alarm is activated by an unstable baseline (for example, leak in the
breathing circuit or at the patient interface). Check for leaks and for improper exhalation valve
operation. If the leak is purposeful (that is, deflated tube cuff), turn NIV on and adjust bias flow to
stabilize PEEP (baseline).
Table 6-2.
Automatic Alarms
High Baseline Pressure
O
2
Cylinder Low/Empty
Low Baseline Pressure
Pressure Control Setting Not Reached
Occlusion
Backup Battery Low Charge
Sustained Occlusion
No External Power
Check Circuit or Prox Line
Shut Down Alert
Device Alert
Backup Battery Low
Power Pac Battery Pack Low
Backup Battery Shutdown Imminent
Integrated Power Pac Failure
Internal Temperature
Switching to Backup Battery
Backup Battery Temperature
Running on Backup Battery
Power Pac Temperature
Motor Fault
Backup Battery Failure
Flow Sensor Disconnect/
Summary of Contents for Newport HT70
Page 1: ...Operator s Manual Newport HT70 Ventilator with Accessories Plus Model TM...
Page 8: ...Page Left Intentionally Blank viii Operator s Manual...
Page 12: ...Page Left Intentionally Blank xii Operator s Manual...
Page 24: ...Introduction Page Left Intentionally Blank 1 12 Operator s Manual...
Page 72: ...Navigating the Screens Page Left Intentionally Blank 4 24 Operator s Manual...
Page 88: ...Operating the Ventilator Page Left Intentionally Blank 5 16 Operator s Manual...
Page 100: ...Ventilator Alarms Page Left Intentionally Blank 6 12 Operator s Manual...
Page 108: ...Battery Operation Page Left Intentionally Blank 7 8 Operator s Manual...
Page 138: ...Explanation of Modes and Controls Page Left Intentionally Blank 10 6 Operator s Manual...
Page 140: ...Quick Check Procedure Check Off Page Left Intentionally Blank A 2 Operator s Manual...
Page 142: ...Foldout Drawings B 2 Page left intentionally blank...
Page 144: ...Foldout Drawings B 4 Page left intentionally blank...
Page 146: ...Foldout Drawings B 6 Page left intentionally blank...
Page 147: ......