B-5
Figure B-3.
HT70PM Model Touch Screen (Hospital Domain)
Note:
While operating on battery power with Power Save enabled and all alarms cleared, the touch screen will go to sleep
after 2 minutes. Just touch the screen to bring it back into view.
1
Screen selections buttons.
Touching any one of these buttons will take you to the new screen. The More screen includes
links to Event, Trends, Wave, and Utility screens.
2
Mode selector.
Touching this button scrolls through the mode choices. The mode will not change until you press the
Accept
button.
3
Breath type selector.
Touching this button toggles the breath type choice. The breath type will not change until you press
the
Accept
button.
4
Help button.
Touching this button enables a tutorial for each feature on the screen. Touch the
Help
button then touch any
button for an explanation of that feature.
5
Monitored data buttons.
Touching any one of these four buttons opens a screen with a view of monitored parameter
choices to display in that button.
6
Message display.
This area shows all informational and alarm messages and current NIV selection, mode, and breath type
selection. During an alarm violation this area will light red for high priority, amber for medium priority, and yellow for low pri-
ority alarms and display the alarm message.
7
Battery charge level indicator.
Shows the charge level of the Power Pac battery pack (blue icon) during external power or
power pac use or the charge level of the backup battery (red icon) during backup battery use.
8
Pressure bar.
Indicates dynamic pressure in the patient circuit in green, the High and Low Pressure alarm settings in red, and
the peak pressure of the last breath in green.
9
Parameter setting buttons.
Touching any one of these buttons will activate the parameter to allow adjustments.
10
Patient effort indicator.
Flashes green to show a spontaneous patient effort.
11
Domain button.
The HT70 can be set up in one of three domains: basic, transport, and hospital. Touch to scroll through the
domain choices. Press
Accept
to confirm choice.
12
AutoLock/unlock button.
This button is only visible if Auto Lock is enabled and the panel is locked. Touch and hold for 3
seconds to unlock touch screen buttons.
Summary of Contents for Newport HT70
Page 1: ...Operator s Manual Newport HT70 Ventilator with Accessories Plus Model TM...
Page 8: ...Page Left Intentionally Blank viii Operator s Manual...
Page 12: ...Page Left Intentionally Blank xii Operator s Manual...
Page 24: ...Introduction Page Left Intentionally Blank 1 12 Operator s Manual...
Page 72: ...Navigating the Screens Page Left Intentionally Blank 4 24 Operator s Manual...
Page 88: ...Operating the Ventilator Page Left Intentionally Blank 5 16 Operator s Manual...
Page 100: ...Ventilator Alarms Page Left Intentionally Blank 6 12 Operator s Manual...
Page 108: ...Battery Operation Page Left Intentionally Blank 7 8 Operator s Manual...
Page 138: ...Explanation of Modes and Controls Page Left Intentionally Blank 10 6 Operator s Manual...
Page 140: ...Quick Check Procedure Check Off Page Left Intentionally Blank A 2 Operator s Manual...
Page 142: ...Foldout Drawings B 2 Page left intentionally blank...
Page 144: ...Foldout Drawings B 4 Page left intentionally blank...
Page 146: ...Foldout Drawings B 6 Page left intentionally blank...
Page 147: ......