Navigating the Screens
4-6
Operator's Manual
WARNING:
Do not use the Newport HT70 Ventilator if the circuit check fails, inadequate ventilation may
result. Use an alternate method of ventilation. Contact Covidien Technical Services.
Note:
The circuit check results are logged into the event history and retained after turning off the ventilator.
4.4.2
Activate Preset Button
While in standby, ventilation and alarm settings can be adjusted manually or they may be adjust-
ed using custom or default preprogrammed parameter sets (presets) for adult, pediatric, and
Infant patients. From the Activate Preset screen, touch the
New Patient Flag
button to enter a “new
patient” flag in the events history log. The ventilator must be in standby condition to activate a
preset or enter a new patient flag.
To use a preset:
1.
Touch the
Activate Preset
button.
Figure 4-4.
Activate Presets
2.
Touch one of the six preset buttons that appear on the screen. Default preset settings are based on
factory set defaults. Custom preset settings must be established by the user.
3.
Press the
Accept
button to implement the settings.
To establish custom parameters for a preset, first set all parameters and alarms as desired for the
patient. Then go to More>Utility>Custom Settings>Set Custom Preset to select a custom preset
button. For instructions on how to customize the custom patient presets, reference
on page
.
Note:
If you do not want to use a preset, press the
Main
,
More,
and
Alarms
buttons and adjust ventilation and
alarm parameters.
Summary of Contents for Newport HT70
Page 1: ...Operator s Manual Newport HT70 Ventilator with Accessories Plus Model TM...
Page 8: ...Page Left Intentionally Blank viii Operator s Manual...
Page 12: ...Page Left Intentionally Blank xii Operator s Manual...
Page 24: ...Introduction Page Left Intentionally Blank 1 12 Operator s Manual...
Page 72: ...Navigating the Screens Page Left Intentionally Blank 4 24 Operator s Manual...
Page 88: ...Operating the Ventilator Page Left Intentionally Blank 5 16 Operator s Manual...
Page 100: ...Ventilator Alarms Page Left Intentionally Blank 6 12 Operator s Manual...
Page 108: ...Battery Operation Page Left Intentionally Blank 7 8 Operator s Manual...
Page 138: ...Explanation of Modes and Controls Page Left Intentionally Blank 10 6 Operator s Manual...
Page 140: ...Quick Check Procedure Check Off Page Left Intentionally Blank A 2 Operator s Manual...
Page 142: ...Foldout Drawings B 2 Page left intentionally blank...
Page 144: ...Foldout Drawings B 4 Page left intentionally blank...
Page 146: ...Foldout Drawings B 6 Page left intentionally blank...
Page 147: ......