Miscellaneous Specifications
Operator's Manual
9-15
WARNING:
The use of accessories and cables other than those specified, with the exception of parts sold by
Covidien as replacements for internal components, may result in increased emissions or decreased
immunity of the ventilator system.
9.11.5
Performance
0.1
0.11 m
0.38 m
0.38 m
0.73 m
1
0.35 m
1.2 m
1.2 m
2.3 m
10
1.1 m
3.8 m
3.8 m
7.3 m
100
3.5 m
12 m
12 m
23 m
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: The ISM (industrial, scientific, and medical) bands between 150 kHz and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553 MHz to
13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.
NOTE 3: An additional factor of 10/3 is used in calculating the recommended separation distance for transmitters in the ISM
frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to 2.5 GHz to decrease the likelihood that
mobile/portable communications equipment could cause interference if it is inadvertently brought into patient areas.
NOTE 4: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
Table 9-20.
Recommended Separation Distances
The ventilator is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the operator of the ventilator can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the ventilator as recommended
below, according to the maximum output power of the communications equipment.
Rated
maximum
output power
of transmitter
(W)
150 kHz to 80 MHz
(outside ISM bands)
150 kHz to 80 MHz
(inside ISM bands)
80MHz to 800 MHz
800 MHz to 2.5 GHz
d
3.5
P
=
d
12
P
=
d
1.2
P
=
d
2.3
P
=
Table 9-21.
Performance Parameter Specifications and Tolerances
Setting
Range
Control accuracy
VT (tidal volume)
50-2200 ml
±10% of setting or ±15mL whichever is greater
Pressure Support
0-60 cmH
2
O
±10% or ±2 cmH
2
O, whichever is greater
Pressure Control
5-60 cmH
2
O
±10% or ±2 cmH
2
O, whichever is greater
Inspiratory time
0.1-3.0 seconds
±0.05 second
Summary of Contents for Newport HT70
Page 1: ...Operator s Manual Newport HT70 Ventilator with Accessories Plus Model TM...
Page 8: ...Page Left Intentionally Blank viii Operator s Manual...
Page 12: ...Page Left Intentionally Blank xii Operator s Manual...
Page 24: ...Introduction Page Left Intentionally Blank 1 12 Operator s Manual...
Page 72: ...Navigating the Screens Page Left Intentionally Blank 4 24 Operator s Manual...
Page 88: ...Operating the Ventilator Page Left Intentionally Blank 5 16 Operator s Manual...
Page 100: ...Ventilator Alarms Page Left Intentionally Blank 6 12 Operator s Manual...
Page 108: ...Battery Operation Page Left Intentionally Blank 7 8 Operator s Manual...
Page 138: ...Explanation of Modes and Controls Page Left Intentionally Blank 10 6 Operator s Manual...
Page 140: ...Quick Check Procedure Check Off Page Left Intentionally Blank A 2 Operator s Manual...
Page 142: ...Foldout Drawings B 2 Page left intentionally blank...
Page 144: ...Foldout Drawings B 4 Page left intentionally blank...
Page 146: ...Foldout Drawings B 6 Page left intentionally blank...
Page 147: ......