Navigating the Screens
4-16
Operator's Manual
Note:
The O
2
monitor (Utility screen) must be enabled for this function to operate.
4.8.5
Calibrate O
2
Mon Screen
The Calibrate O
2
Mon screen is available in the hospital and transport domains and in ventilating
condition.
Figure 4-12.
Calibrate O
2
Mon Screen
This screen allows the user to calibrate the internal oxygen sensor. Either a single-point or a two-
point calibration can be done. This can be done while on a patient if they can tolerate the desired
calibration point (room air or 100% oxygen). Touch this button to go to the Calibrate O
2
Mon
screen.
O
2
Cal, 21% O
2
: Touch this button, and follow the on-screen directions to calibrate at room air.
Ensure that no oxygen device is connected to the air intake port on the right side of the ventilator.
O
2
Cal, 100% O
2
: Touch this button, and follow the on-screen directions to calibrate at 100% ox-
ygen. Ensure that 100% oxygen is being delivered to the air intake port on the right side of the
ventilator. Covidien suggests using the low flow oxygen reservoir with 10 L/min of medical-grade
100% oxygen connected to it.
4.9
Utility Screen
The Utility screen is available in the hospital and transport domains and in standby and ventilating
condition except where noted.
Access the Utility screen through the More screen, as described in
.
Reference
on page
Summary of Contents for Newport HT70
Page 1: ...Operator s Manual Newport HT70 Ventilator with Accessories Plus Model TM...
Page 8: ...Page Left Intentionally Blank viii Operator s Manual...
Page 12: ...Page Left Intentionally Blank xii Operator s Manual...
Page 24: ...Introduction Page Left Intentionally Blank 1 12 Operator s Manual...
Page 72: ...Navigating the Screens Page Left Intentionally Blank 4 24 Operator s Manual...
Page 88: ...Operating the Ventilator Page Left Intentionally Blank 5 16 Operator s Manual...
Page 100: ...Ventilator Alarms Page Left Intentionally Blank 6 12 Operator s Manual...
Page 108: ...Battery Operation Page Left Intentionally Blank 7 8 Operator s Manual...
Page 138: ...Explanation of Modes and Controls Page Left Intentionally Blank 10 6 Operator s Manual...
Page 140: ...Quick Check Procedure Check Off Page Left Intentionally Blank A 2 Operator s Manual...
Page 142: ...Foldout Drawings B 2 Page left intentionally blank...
Page 144: ...Foldout Drawings B 4 Page left intentionally blank...
Page 146: ...Foldout Drawings B 6 Page left intentionally blank...
Page 147: ......