Alarms Screen Navigation
Operator's Manual
4-7
4.4.3
Start Ventilation Button
Touch the
Start Ventilation
button to exit the standby condition and begin ventilation.
WARNING:
Ensure that all settings are appropriate for the patient prior to starting ventilation. Note that
during standby condition, the monitored O
2
is not representative of the set O
2
or the O
2
that will
be delivered during ventilation. After starting ventilation, use a calibrated oxygen monitor (such
as the one that is built-in to the ventilator) to verify that the air oxygen entrainment mixer setting
or liter flow attached to the low flow oxygen reservoir is delivering the prescribed O
2
.
Note:
Be sure to review all chapters of this manual before you use the Newport HT70 Ventilator for the first time.
4.5
Alarms Screen Navigation
The Alarms screen can be accessed in all domains and in standby or ventilating condition except
where noted.
on page
for each alarm and setting.
Enter the Alarms Screen
Touch the
Alarms
button.
Set Alarm Limits (Not Available in Basic Domain)
Touch an
Alarm limit
button to activate (highlight) it, and then use the up and down arrow panel
buttons to adjust the limit. Press
Accept
to confirm changes or press
Cancel
to return to original
settings. Multiple limits can be set before pressing
Accept
.
Summary of Contents for Newport HT70
Page 1: ...Operator s Manual Newport HT70 Ventilator with Accessories Plus Model TM...
Page 8: ...Page Left Intentionally Blank viii Operator s Manual...
Page 12: ...Page Left Intentionally Blank xii Operator s Manual...
Page 24: ...Introduction Page Left Intentionally Blank 1 12 Operator s Manual...
Page 72: ...Navigating the Screens Page Left Intentionally Blank 4 24 Operator s Manual...
Page 88: ...Operating the Ventilator Page Left Intentionally Blank 5 16 Operator s Manual...
Page 100: ...Ventilator Alarms Page Left Intentionally Blank 6 12 Operator s Manual...
Page 108: ...Battery Operation Page Left Intentionally Blank 7 8 Operator s Manual...
Page 138: ...Explanation of Modes and Controls Page Left Intentionally Blank 10 6 Operator s Manual...
Page 140: ...Quick Check Procedure Check Off Page Left Intentionally Blank A 2 Operator s Manual...
Page 142: ...Foldout Drawings B 2 Page left intentionally blank...
Page 144: ...Foldout Drawings B 4 Page left intentionally blank...
Page 146: ...Foldout Drawings B 6 Page left intentionally blank...
Page 147: ......