Domain Navigation
Operator's Manual
4-21
The BUV Settings screen includes the following adjustable parameters:
•
Minimum RR (respiratory rate)
•
Rate Factor (set rate will be multiplied times this number to determine the BUV breath rate)
•
SPONT (mode) delta P (pressure target above set PEEP for breath delivery)
•
SPONT (mode) i-Time (i-time for BUV breaths delivered while in SPONT mode)
•
BUV link (with LMV [low minute volume] alarm, Apnea alarm, or both)
•
Revert to Defaults
Press the
Accept
button to confirm changes made to any of the parameters.
4.11
Domain Navigation
The Newport HT70 Ventilator is designed with the flexibility of useful application in the acute care
as well as long-term care environments. To make the product easy and safe to use in the full spec-
trum of applications, care is divided into three domains: hospital (which means acute care), trans-
port (anytime the user is on the move with battery and supplemental oxygen supplies), and basic
(for the long-term/homecare environments).
Touch the
Domain
button in the lower right corner of the touch screen to scroll through the
domain choices: Hosp (hospital), Trans (transport), and Basic (basic). Press
Accept
to change to the
new domain. The
Domain
button is not visible if the panel is locked.
4.11.1
Hospital Domain
The hospital domain has full access to all features and screen selections available on the ventilator.
This manual describes all of the features and screen selections found in the hospital domain. The
transport and basic domain features and screen selections work identically to the hospital domain
with the limitations as noted in their respective sections.
Summary of Contents for Newport HT70
Page 1: ...Operator s Manual Newport HT70 Ventilator with Accessories Plus Model TM...
Page 8: ...Page Left Intentionally Blank viii Operator s Manual...
Page 12: ...Page Left Intentionally Blank xii Operator s Manual...
Page 24: ...Introduction Page Left Intentionally Blank 1 12 Operator s Manual...
Page 72: ...Navigating the Screens Page Left Intentionally Blank 4 24 Operator s Manual...
Page 88: ...Operating the Ventilator Page Left Intentionally Blank 5 16 Operator s Manual...
Page 100: ...Ventilator Alarms Page Left Intentionally Blank 6 12 Operator s Manual...
Page 108: ...Battery Operation Page Left Intentionally Blank 7 8 Operator s Manual...
Page 138: ...Explanation of Modes and Controls Page Left Intentionally Blank 10 6 Operator s Manual...
Page 140: ...Quick Check Procedure Check Off Page Left Intentionally Blank A 2 Operator s Manual...
Page 142: ...Foldout Drawings B 2 Page left intentionally blank...
Page 144: ...Foldout Drawings B 4 Page left intentionally blank...
Page 146: ...Foldout Drawings B 6 Page left intentionally blank...
Page 147: ......