Miscellaneous Specifications
Operator's Manual
9-13
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC/EN 61000-4-11
<5% U
T
(>95% dip in U
T
) for 0.5
cycle
40% U
T
(60% dip in U
T
) for 5
cycles
70% U
T
(30% dip in U
T
) for 25
cycles
<5% U
T
>95% dip in U
T
for 5 s
<5% U
T
(>95% dip in U
T
) for 0.5
cycle
40% U
T
(60% dip in U
T
) for 5
cycles
70% U
T
(30% dip in U
T
) for 25
cycles
<5% U
T
>95% dip in U
T
for 5 s
AC power (“mains”) power quality
should be that of a typical
commercial or hospital environment.
If the user of the ventilator requires
continued operation during power
mains interruptions, it is
recommended that the ventilator be
powered from an uninterruptible
power supply or a battery.
Power frequency (50/60 Hz)
magnetic field
IEC/EN 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical
commercial or hospital environment.
NOTE: U
T
is the AC mains voltage prior to application of the test level.
Table 9-19.
Electromagnetic Immunity—Conducted and Radiated RF
The ventilator is intended for use in the electromagnetic environment specified below. The customer or the operator of
the ventilator should assure that it is used in such an environment.
Immunity test
IEC/EN 60601 test level
Compliance level
Electromagnetic environment—
guidance
Conducted RF
IEC/EN 61000-4-6
3 Vrms
150 kHz to 80 MHz
outside ISM bands
1
3 Vrms
150 kHz to 80 MHz
outside ISM bands
1
Portable and mobile RF communications
equipment should be used no closer to any
part of the ventilator system, including
cables, than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance:
10 Vrms
inside ISM bands
1
10 Vrms
inside ISM bands
1
Recommended separation distance:
Table 9-18.
Electromagnetic Immunity (Continued)
The ventilator is intended for use in the electromagnetic environment specified below. The customer or the user of the
ventilator should assure that it is used in such an environment.
Immunity test
IEC/EN 60601 test level
Compliance level
Electromagnetic environment—
guidance
d
0.35
P
=
d
1.2
P
=
Summary of Contents for Newport HT70
Page 1: ...Operator s Manual Newport HT70 Ventilator with Accessories Plus Model TM...
Page 8: ...Page Left Intentionally Blank viii Operator s Manual...
Page 12: ...Page Left Intentionally Blank xii Operator s Manual...
Page 24: ...Introduction Page Left Intentionally Blank 1 12 Operator s Manual...
Page 72: ...Navigating the Screens Page Left Intentionally Blank 4 24 Operator s Manual...
Page 88: ...Operating the Ventilator Page Left Intentionally Blank 5 16 Operator s Manual...
Page 100: ...Ventilator Alarms Page Left Intentionally Blank 6 12 Operator s Manual...
Page 108: ...Battery Operation Page Left Intentionally Blank 7 8 Operator s Manual...
Page 138: ...Explanation of Modes and Controls Page Left Intentionally Blank 10 6 Operator s Manual...
Page 140: ...Quick Check Procedure Check Off Page Left Intentionally Blank A 2 Operator s Manual...
Page 142: ...Foldout Drawings B 2 Page left intentionally blank...
Page 144: ...Foldout Drawings B 4 Page left intentionally blank...
Page 146: ...Foldout Drawings B 6 Page left intentionally blank...
Page 147: ......