Clinical Studies
A-36
Operator’s Manual
The results are presented in Table A-11, showing an increase in percent modulation
greater than 2.5-fold during motion.
Pulse oximeters have been known to perform better at the higher saturation levels
compared to the lower end. Though, when presenting the A
RMS
, the common
methodology is to provide the data across the whole range (70% to 100%). The data
below is presented to show each decade, which includes the RMSD and N values.
There is no acceptance criteria associated with decade levels of hypoxia. In Table A-
12, SpO
2
RMSD is presented per decade.
RMSD for each decade are well within the acceptance criteria of 3%. The plateaus
that were used during the study were 70 - 76, 76.01 - 85, 85.01 - 94 and >94%, as pre-
sented in Table A-13.
Table A-11.
Percent Modulation During Motion
Quiescent
Periods
Motion
Periods
Ratio
Percent
Modulation (%)
1.63
4.14
2.54
Table A-12.
RMSD of SpO
2
per Decade (Motion)
SpO
2
Range
100%-90%
89%-80%
79%-70%
N
589
322
344
RMSD (%)
2.36
1.97
2.41
Summary of Contents for Nellcor Series
Page 1: ...Operator s Manual Nellcor Bedside SpO2 Patient Monitoring System TM...
Page 8: ...Page Left Intentionally Blank vi...
Page 12: ...Page Left Intentionally Blank x...
Page 20: ...Introduction Page Left Intentionally Blank 1 8 Operator s Manual...
Page 36: ...Installation Page Left Intentionally Blank 3 8 Operator s Manual...
Page 118: ...Product Specifications Page Left Intentionally Blank 11 18 Operator s Manual...
Page 156: ...Clinical Studies Page Left Intentionally Blank A 38 Operator s Manual...
Page 159: ......