Product Specifications
11-10
Operator’s Manual
Electromagnetic Immunity
Table 11-7.
Electromagnetic Immunity Guidelines and Compliance
Guidance and Manufacturer’s Declaration—Electromagnetic Immunity
(IEC/EN 60601-1-2)
The monitoring system is intended for use in the electromagnetic environment specified below. The cus-
tomer or the user of the monitoring system should assure that it is used in such an environment.
Immunity
Test
IEC/EN 60601-1-2
Test Level
Compliance
Level
Electromagnetic Environment
Guidance
Electrostatic
discharge (ESD)
IEC/EN 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
Floor should be wood, concrete, or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Electric fast
transient/burst
IEC/EN 61000-4-4
± 2 kV for power
supply lines
± 1 kV input/output
lines
± 2 kV for
power supply lines
± 1 kV input/output
lines
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC/EN 61000-4-5
± 1 kV differential
mode
± 2 kV common
mode
± 1 kV differential
mode
± 2 kV common
mode
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips and
interrupts
IEC/EN 61000-4-11
100% reduction for
0.5 cycles (at 0°, 45°,
90°, 135°, 180°, 225°,
270°, and 315°)
100% reduction for
1.0 cycle (at 0°)
30% reduction for
25/30 cycles (at 0°)
100% reduction for
250/300 cycles (at
0°)
100% reduction for
0.5 cycles (at 0°, 45°,
90°, 135°, 180°, 225°,
270°, and 315°)
100% reduction for
1.0 cycle (at 0°)
30% reduction for
25/30 cycles (at 0°)
100% reduction for
250/300 cycles (at
0°)
Mains power quality should be that
of a typical commercial or hospital
environment.
If the user requires continued oper-
ation during power mains interrup-
tion, it is recommended that the
monitoring system be powered
from an uninterruptible power
supply or battery.
Power frequency (50/
60 Hz) magnetic field
IEC/EN 61000-4-8
30 A/m
30 A/m
It may be necessary to position
further from the sources of power
frequency magnetic fields or to
install magnetic shielding.
Summary of Contents for Nellcor Series
Page 1: ...Operator s Manual Nellcor Bedside SpO2 Patient Monitoring System TM...
Page 8: ...Page Left Intentionally Blank vi...
Page 12: ...Page Left Intentionally Blank x...
Page 20: ...Introduction Page Left Intentionally Blank 1 8 Operator s Manual...
Page 36: ...Installation Page Left Intentionally Blank 3 8 Operator s Manual...
Page 118: ...Product Specifications Page Left Intentionally Blank 11 18 Operator s Manual...
Page 156: ...Clinical Studies Page Left Intentionally Blank A 38 Operator s Manual...
Page 159: ......