![background image](http://html1.mh-extra.com/html/covidien/nellcor-series/nellcor-series_operators-manual_2673906086.webp)
Troubleshooting
8-2
Operator’s Manual
8.3
Error Conditions
Table 8-1.
Common Problems and Resolutions
Problem
Resolution
Battery Charging Indicator not lit
Check power cord
Check battery
Check AC power inlet
Check power/ mains outlet
Sensor Message
SpO
2
Pulse search
Patient Movement Detected
SpO
2
Sensor Off
SpO
2
Cable/Sensor Disconnect
SpO
2
Loss of Pulse
Check patient status; keep patient still, check for perfusion
Check all connections
Reposition sensor
Check or change adhesive wrap
Choose alternate site
Warm site
Cover sensor
Use forehead, nasal, or ear sensor (adult patient only)
Use Nellcor™ adhesive sensor
Secure cable
Secure with headband (MAXFAST)
Remove nail polish
Loosen sensor (too tight)
Isolate external interference (electrosurgical device, cell phone)
Replace the cable and/or sensor
Clean site (MAXR)
No response to Power On/Off
button press
Press the Power On/Off button for more than one (1) second.
Ensure the power cord is properly connected to the outlet.
Ensure AC indicator blinks.
Ensure it does not share the same AC power source with other equipment.
If the error continues, contact Technical Services or a qualified service technician.
No response to button press
Verify whether the Home button has not been pressed during normal screen.
If the error continues, contact Technical Services or a qualified service technician.
Frozen at POST after power on
Power cycle by pressing the Power On/Off button.
If the error continues, contact Technical Services or a qualified service technician.
System is frozen
If the system freezes, it generates beep tone. Press the power button over 15
seconds for force quit the system.
If the error continues, contact Technical Services or a qualified service technician.
Summary of Contents for Nellcor Series
Page 1: ...Operator s Manual Nellcor Bedside SpO2 Patient Monitoring System TM...
Page 8: ...Page Left Intentionally Blank vi...
Page 12: ...Page Left Intentionally Blank x...
Page 20: ...Introduction Page Left Intentionally Blank 1 8 Operator s Manual...
Page 36: ...Installation Page Left Intentionally Blank 3 8 Operator s Manual...
Page 118: ...Product Specifications Page Left Intentionally Blank 11 18 Operator s Manual...
Page 156: ...Clinical Studies Page Left Intentionally Blank A 38 Operator s Manual...
Page 159: ......