Operation
4-2
Operator’s Manual
WARNING:
Do not use damaged pulse oximetry sensors. Do not use with exposed optical
components. Do not immerse completely in water, solvents, or cleaning solutions,
since pulse oximetry sensors and connectors are not waterproof. Do not sterilize by
irradiation, steam or ethylene oxide. Refer to the cleaning instructions in the
Directions for Use
for reusable sensors.
Caution:
Do not attach any cable intended for computer use to the sensor port connector.
Caution:
The sensor disconnect error message and associated alarm indicate the pulse
oximetry sensor is either disconnected or has faulty wiring. Check the connection
and, if necessary, replace it, the pulse oximetry cable, or both.
4.3
Apply Power to Monitoring System
4.3.1
Turn On the Monitoring System
WARNING:
Ensure the speaker is clear of any obstruction. Failure to do so may result in an
inaudible tone.
Caution:
If any indicator or display element does not light, or the speaker does not sound, do not
use the monitoring system. Instead, contact a qualified service technician.
To power on the monitoring system:
1.
Press the
Power On/Off Button
for more than one (1) second.
2.
Ensure the software version, the SpO
2
alarm indicator, and the pulse rate alarm indicator
light for approximately two (2) seconds.
Summary of Contents for Nellcor Series
Page 1: ...Operator s Manual Nellcor Bedside SpO2 Patient Monitoring System TM...
Page 8: ...Page Left Intentionally Blank vi...
Page 12: ...Page Left Intentionally Blank x...
Page 20: ...Introduction Page Left Intentionally Blank 1 8 Operator s Manual...
Page 36: ...Installation Page Left Intentionally Blank 3 8 Operator s Manual...
Page 118: ...Product Specifications Page Left Intentionally Blank 11 18 Operator s Manual...
Page 156: ...Clinical Studies Page Left Intentionally Blank A 38 Operator s Manual...
Page 159: ......