Choose Patient Settings
Operator’s Manual
4-5
Figure 4-2.
Save Change Screen
The LCD panel displays easy-to-read numeric values for patient oxygen saturation
and pulse rate in cyan and green, respectively. Reference
4.5
Choose Patient Settings
4.5.1
Set Patient Mode (Type)
Select the desired patient mode: Adult, Pediatric or Neonatal.
To set patient mode:
1.
Select the Patient Mode icon.
2.
Select the correct patient mode (type):
Adult: Use for adults.
Pediatric: Use for children.
Neonatal: Use for newborns.
Note:
Use the appropriate patient mode and pulse oximetry sensor based on body weight.
Reference the pulse oximetry sensor
Instructions for Use
.
4.5.2
Set SpO
2
and Pulse Rate Alarm Limits
WARNING:
Do not silence the audible alarm or decrease its volume if patient safety could be
compromised.