Results
Operator’s Manual
A-23
The plateaus that were used during the study were 70 - 76, 76.01 - 85, 85.01 - 94 and
>94%, as presented in Table A-4.
A.3.2
Accuracy Results (Low Saturation)
The accuracy results for both SpO
2
and pulse rate can be found in Table A-5 and
Table A-6 across a SaO
2
range of 60 to 80%. (Reference
2
range 70 to 100%.) As shown in the tables, both SpO
2
and
pulse rate meet the acceptance criteria for the MAXA, MAXN, and MAXFAST sensors
during non-motion conditions.
Table A-4.
RMSD of SpO
2
per Plateau (No Motion)
SpO
2
Range
Room Air
Plateau
90%
Plateau
80%
Plateau
70%
Plateau
N
978
1102
1034
828
RMSD (%)
1.27
1.65
1.69
2.15
Table A-5.
SpO
2
Accuracy Results (60 to 80% SaO
2
)
Sensor
Motion
# of
Data
Points
A
RMS
(%)
SpO
2
Acceptance
Criteria
60%-80%
(%)
Pass/Fail
MAXA
No
610
2.40
3.0
Pass
MAXN
No
453
1.92
3.0
Pass
MAXFAST
No
143
2.41
3.0
Pass
ALL
No
1206
2.24
3.0
Pass
Table A-6.
Pulse Rate Accuracy Results (60 to 80% SaO
2
)
Sensor
Motion
# of
Data
Points
A
RMS
(BPM)
Pulse Rate
Acceptance
Criteria
(BPM)
Pass/Fail
MAXA
No
1154
0.76
3.0
Pass
MAXN
No
874
0.74
3.0
Pass
Summary of Contents for Nellcor Series
Page 1: ...Operator s Manual Nellcor Bedside SpO2 Patient Monitoring System TM...
Page 8: ...Page Left Intentionally Blank vi...
Page 12: ...Page Left Intentionally Blank x...
Page 20: ...Introduction Page Left Intentionally Blank 1 8 Operator s Manual...
Page 36: ...Installation Page Left Intentionally Blank 3 8 Operator s Manual...
Page 118: ...Product Specifications Page Left Intentionally Blank 11 18 Operator s Manual...
Page 156: ...Clinical Studies Page Left Intentionally Blank A 38 Operator s Manual...
Page 159: ......