Safety Information
Operator’s Manual
1-3
WARNING:
Always disconnect and remove the monitoring system and sensors during magnetic
resonance imaging (MRI) scanning. Attempting to use the monitoring system during
an MRI procedure could cause burns or adversely affect the MRI image or the
monitoring system's accuracy.
WARNING:
The monitoring system is intended only as an adjunct in patient assessment. It must
be used in conjunction with clinical signs and symptoms.
WARNING:
The values displayed by the monitoring system can be affected by patient
conditions, excessive patient movement, sensors, environmental conditions, and
nearby external electromagnetic conditions.
WARNING:
When the monitoring system is set to Standard Mode, it is intended for use in a
hospital or hospital-type environment by trained medical personnel.
WARNING:
Failure to cover the pulse oximetry sensor site with opaque material in high ambient
light conditions may result in inaccurate measurements. Refer to the appropriate
sections of this manual for specific safety information.
WARNING:
The monitoring system may remain attached to the patient during defibrillation or
during use of an electrosurgical unit; however, the monitoring system is not
defibrillator-proof, and readings may be inaccurate during defibrillation and shortly
thereafter.
WARNING:
The monitoring system may retain trend data from multiple patients if transferred
from one patient to another.
WARNING:
Any connections between this monitoring system and other devices must comply
with applicable medical systems safety standards such as IEC 60601-1. Failure to do
so could result in unsafe leakage current and grounding conditions.
Summary of Contents for Nellcor Series
Page 1: ...Operator s Manual Nellcor Bedside SpO2 Patient Monitoring System TM...
Page 8: ...Page Left Intentionally Blank vi...
Page 12: ...Page Left Intentionally Blank x...
Page 20: ...Introduction Page Left Intentionally Blank 1 8 Operator s Manual...
Page 36: ...Installation Page Left Intentionally Blank 3 8 Operator s Manual...
Page 118: ...Product Specifications Page Left Intentionally Blank 11 18 Operator s Manual...
Page 156: ...Clinical Studies Page Left Intentionally Blank A 38 Operator s Manual...
Page 159: ......