Operation
4-20
Operator’s Manual
4.9
Menu Structure and Factory Defaults
The monitoring system ships with factory default settings. To set different institu-
tional default settings, contact a qualified service technician.
Note:
The ability to change alarm limits is inhibited in Homecare Mode.
Table 4-3.
Menu Structure and Factory Defaults
Item
Available Selections
Factory Default
Adult
Pediatric
Neonatal
QUICK ACCESS ALARM LIMITS Menus and ALARM/LIMIT Menu
SpO
2
Menu
Upper SpO
2
alarm limit (21-100)
100%
95%
Lower SpO
2
alarm limit (20-99)
90%
85%
Alarm Inhibition for SpO
2
alarms
Off
SatSeconds™ alarm management setting (Off, 10,
25, 50, 100)
Off
Pulse Rate (PR) Menu
Upper pulse rate alarm limit (30-245)
5 bpm steps
170 bpm
200 bpm
Lower pulse rate alarm limit (25-240)
5 bpm steps
50 bpm
75 bpm
100 bpm
Alarm Inhibition for pulse rate alarms
Off
OPTIONS Menu
Volume
Alarm Volume (1-8)
5
Key Beep Volume (Off, 1-7)
4
Pulse Volume (Off, 1-7)
4
Brightness
Screen brightness (1-8)
4
Change Patient Mode
Mode—Response Mode (Normal, Fast)
Normal
Monitoring Mode (Standard, Homecare, Sleep
Study)
Standard Mode
Switch to Homecare Mode
--
Switch to Sleep Study Mode
--
Transfer Data
Continuous Transfer, Trend Data Download
By USB
Summary of Contents for Nellcor Series
Page 1: ...Operator s Manual Nellcor Bedside SpO2 Patient Monitoring System TM...
Page 8: ...Page Left Intentionally Blank vi...
Page 12: ...Page Left Intentionally Blank x...
Page 20: ...Introduction Page Left Intentionally Blank 1 8 Operator s Manual...
Page 36: ...Installation Page Left Intentionally Blank 3 8 Operator s Manual...
Page 118: ...Product Specifications Page Left Intentionally Blank 11 18 Operator s Manual...
Page 156: ...Clinical Studies Page Left Intentionally Blank A 38 Operator s Manual...
Page 159: ......