Data Management
5-10
Operator’s Manual
2.
Ensure patient data settings coincide with expected settings. This would include the
version of firmware and its CRC code, which should be all zeros; the current method of
viewing the data: waveform, trend, or graph; alarm limit settings; patient mode; and
SatSeconds™ setting.
3.
Scan the time, SpO
2
, or PR column until reaching the events of interest.
4.
Match the operating status codes to the following table for pertinent system informa-
tion. Reference
COM port USB Driver Alternatives
•
Load the appropriate driver from the product CD or thumb drive into the connected
computer. The USB driver will automatically load.
•
Contact Technical Services or a local Covidien representative.
6
PR
Current pulse rate
7
PA
Current pulse ampltitude
8
Status
Operating status of the monitoring system
Table 5-3.
Status Codes
Code
Definition
Code
Definition
AO
Alarm off
PH
Pulse rate upper limit alarm
AS
Alarm silence
PL
Pulse rate lower limit alarm
BU
Battery in use
PS
Pulse search
LB
Low battery
SD
Sensor disconnect
LM
Loss of pulse with patient motion
SH
Saturation rate upper limit alarm
LP
Loss of pulse
SL
Saturation rate lower limit alarm
ID
Patient motion detected
SO
Sensor off
MO
Patient motion
Summary of Contents for Nellcor Series
Page 1: ...Operator s Manual Nellcor Bedside SpO2 Patient Monitoring System TM...
Page 8: ...Page Left Intentionally Blank vi...
Page 12: ...Page Left Intentionally Blank x...
Page 20: ...Introduction Page Left Intentionally Blank 1 8 Operator s Manual...
Page 36: ...Installation Page Left Intentionally Blank 3 8 Operator s Manual...
Page 118: ...Product Specifications Page Left Intentionally Blank 11 18 Operator s Manual...
Page 156: ...Clinical Studies Page Left Intentionally Blank A 38 Operator s Manual...
Page 159: ......