![background image](http://html1.mh-extra.com/html/covidien/nellcor-series/nellcor-series_operators-manual_2673906113.webp)
Manufacturer’s Declaration and Guidance
Operator’s Manual
11-13
Table 11-10.
Test Specifications for Enclosure Port Immunity to RF Wireless Communications Equipment
Test
frequency
(MHz)
Band
(MHz)
Service
Modulation
Max. power
(W)
Distance
(m)
Immunity
test level
(V/m)
385
380 to 390
TETRA 400
Pulse modulation
18 Hz
1.8
0.3
27
450
430 to 470
GMRS 460, FRS 460
FM
± 5kHz deviation
1 kHz sine
2
0.3
28
710
704 to 787
LTE Band 13, 17
Pulse modulation
217 Hz
0.2
0.3
9
745
780
810
800 to 960
GSM 800/900, TETRA
800, iDEN 820, CDMA
850, LTE Band 5
Pulse modulation
18 Hz
2
0.3
28
870
930
1720
1700 to
1990
GSM 1800; CDMA
1900; GSM 1900;
DECT; LTE Band 1, 3,
4, 25; UMTS
Pulse modulation
217 Hz
2
0.3
28
1845
1970
2450
2400 to
2570
Bluetooth, WLAN,
802.11 b/g/n, RFID
2450, LTE Band 7
Pulse modulation
217 Hz
2
0.3
28
5240
5100 to
5800
WLAN 802.11 a/n
Pulse modulation
217 Hz
0.2
0.3
9
5500
5785
Summary of Contents for Nellcor Series
Page 1: ...Operator s Manual Nellcor Bedside SpO2 Patient Monitoring System TM...
Page 8: ...Page Left Intentionally Blank vi...
Page 12: ...Page Left Intentionally Blank x...
Page 20: ...Introduction Page Left Intentionally Blank 1 8 Operator s Manual...
Page 36: ...Installation Page Left Intentionally Blank 3 8 Operator s Manual...
Page 118: ...Product Specifications Page Left Intentionally Blank 11 18 Operator s Manual...
Page 156: ...Clinical Studies Page Left Intentionally Blank A 38 Operator s Manual...
Page 159: ......