Data Management
5-8
Operator’s Manual
9.
Wait for the
Output Complete
message to indicate the download is complete.
10.
Save patient trend data to the personal computer disk or to an alternate source,
depending on institutional requirements.
To launch HyperTerminal
1.
Click the
Start
menu in the main task bar.
2.
Mouse over the
Programs
submenu, then
Accessories
, then
Communications
, then
the
HyperTerminal
option.
Note:
If this is the first time the HyperTerminal program launches, it will prompt the user to set
it as the default Telnet program. Depending on institutional requirements, choose Yes
or No.
3.
Click the HyperTerminal option.
4.
When the Connect Description window opens, type in the desired file name in the
Name field.
5.
Locate the proper icon by scrolling all the way to the far right of the icon field.
6.
Select the icon.
7.
Click the OK button.
Note:
If the personal computer is not connected via the USB to mini-USB cable to the
monitoring system, the proper COM port option will not appear in the list.
8.
When the Connect To window opens, find the
Connect Using
option and click the
down arrow to identify possible modem options.
9.
Select the desired COM port.
10.
Click the OK button.
11.
In the COM Properties window, set the appropriate values.
a.
Set the baud rate (bits per second) to match the monitoring system. The factory
default baud rate is 19200 bits per second (bps).
b.
Ensure the data bit is set to 8.
Summary of Contents for Nellcor Series
Page 1: ...Operator s Manual Nellcor Bedside SpO2 Patient Monitoring System TM...
Page 8: ...Page Left Intentionally Blank vi...
Page 12: ...Page Left Intentionally Blank x...
Page 20: ...Introduction Page Left Intentionally Blank 1 8 Operator s Manual...
Page 36: ...Installation Page Left Intentionally Blank 3 8 Operator s Manual...
Page 118: ...Product Specifications Page Left Intentionally Blank 11 18 Operator s Manual...
Page 156: ...Clinical Studies Page Left Intentionally Blank A 38 Operator s Manual...
Page 159: ......