Set Additional Patient Modes
Operator’s Manual
4-17
Note:
It is recommended to delete all trend data when configuring the monitoring system for
a different patient.
The “Switched to Sleep Study Mode” message appears.
Figure 4-17.
Sleep Study Mode Main Screen
When the buttons and knob have not been used for 30 seconds, the screen dims.
4.7.4
Standard Mode
Standard Mode is the factory default operating mode. If Homecare Mode or Sleep
Study Mode is the current setting, set the monitoring system back to Standard Mode
when a clinician will be using it in a hospital or other professional care setting.
Note:
A four-digit password is required to switch to Standard Mode.
To set the monitoring system to Standard Mode:
1.
Access the OPTIONS Menu.
2.
Select
Change Patient Mode
, then select
Switch to Standard Mode
.
3.
Enter the password for Standard Mode.
4.
When prompted, confirm or cancel the switch to Standard Mode.
5.
Choose whether to keep or delete all trend data.