SatSeconds™ Alarm Management Feature
Operator’s Manual
10-5
10.6.1
First SpO
2
Event
Consider the first event. Suppose the SatSeconds™ alarm limit is set to 25. The
patient’s SpO
2
drops to 79% and the duration of the event is two (2) seconds before
saturation again exceeds the lower alarm threshold of 85%.
Because the SatSeconds™ alarm limit is set to 25 and the actual number of SatSec-
onds™ equals 12, there is no audible alarm.
Figure 10-3.
First SpO
2
Event: No SatSeconds™ Alarm
6% drop below the lower alarm limit threshold
x 2 second duration below the lower threshold
12 SatSeconds™
; no alarm