REF C04717-02-91 Q-NRG User Manual, Edition 4.2
Chapter 5 - Maintenance
COSMED srl
Syringe Adapter
Level of Reprocessing
Low-level disinfection
When
Immediately after use
Pre-treatment
Wipe soil with a moist sponge or towel
Manual Cleaning
Wipe device surface with a soft cloth sprayed with cleaning solution (e.g. Oxyvir® Tb RTU).
Dispose the soft cloth.
Disinfection
Repeat manual cleaning. Wipe device surface with a soft cloth moistened with water. Dispose
the soft cloth.
Drying
Air dry
Material Facts
EVA copolymer
Corrugated tube for adapter
Level of Reprocessing
Low-level disinfection
When
Immediately after use
Pre-treatment
Wipe soil with a moist sponge or towel
Manual Cleaning
Wipe device surface with a soft cloth sprayed with cleaning solution (e.g. Oxyvir® Tb RTU).
Dispose the soft cloth.
Disinfection
Repeat manual cleaning. Wipe device surface with a soft cloth moistened with water. Dispose
the soft cloth.
Drying
Air dry
Material Facts
ABS plastic, mesh
Note:
Components which have been disassembled have to be reassembled after reprocessing.
Summary of Contents for Q-NRG
Page 1: ...Q NRG User Manual...
Page 2: ......
Page 32: ...30 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Page 40: ...38 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Page 86: ...84 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Page 121: ...119 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl Appendix...
Page 148: ...146 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl Notes...
Page 149: ...147 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl...
Page 150: ...148 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl...
Page 151: ......